Home A1c tests: how accurate are they?

An at-home A1c kit uses a drop of blood from a fingerstick. It can show roughly where you are between lab tests. Here is how far a kit result can sit from a lab result, and what a kit cannot do.

Last reviewed: By the controla1clevels.com editorial team General information, not medical advice.

The short answer

What "accurate" means in numbers

Accuracy rules for A1c are written as a percent of the true value, not as A1c points. Here is what each yardstick allows around a lab result:

Lab A1cWithin 5%Within 8%Within 13.5%
6.5% 6.2% to 6.8%6.0% to 7.0%5.6% to 7.4%
7.0% 6.7% to 7.4%6.4% to 7.6%6.1% to 7.9%
8.0% 7.6% to 8.4%7.4% to 8.6%6.9% to 9.1%
9.0% 8.6% to 9.5%8.3% to 9.7%7.8% to 10.2%

Simple arithmetic: lab value × (1 ± percent), rounded to one decimal. Sources: NGSP criteria (NGSP criteria), NGSP on proficiency testing (NGSP news), Chang 2010 (Chang 2010).

So with a true A1c of 7.0%, a home result anywhere from 6.1% to 7.9% would have counted as accurate in that study. That spread crosses several key values, which is why a home result is a guide, not a verdict.

What the studies found

Devices differ, and so do lots of the same device: the Dutch study found significant differences between reagent lots for every point-of-care instrument it tested (Lenters-Westra 2010).

How to check a kit

When a home result and a lab result disagree

Got a home result? See what it means in average glucose with the A1c converter. Retesting soon after a change? Read how fast A1c can change.

Sources

  1. American Diabetes Association Professional Practice Committee. 2. Diagnosis and Classification of Diabetes: Standards of Care in Diabetes 2026. Diabetes Care. 2026;49(Suppl 1). doi.org/10.2337/dc26-S002
  2. Chang A, Frank J, Knaebel J, Fullam J, Pardo S, Simmons DA. Evaluation of an over-the-counter glycated hemoglobin (A1C) test kit. J Diabetes Sci Technol. 2010;4(6):1495-1503. doi.org/10.1177/193229681000400625
  3. Lenters-Westra E, Slingerland RJ. Six of eight hemoglobin A1c point-of-care instruments do not meet the general accepted analytical performance criteria. Clin Chem. 2010;56(1):44-52. doi.org/10.1373/clinchem.2009.130641
  4. Nathan DM, Griffin A, Perez FM, Basque E, Do L, Steiner B. Accuracy of a point-of-care hemoglobin A1c assay. J Diabetes Sci Technol. 2019;13(6):1149-1153. doi.org/10.1177/1932296819836101
  5. NGSP. Summary of certification criteria (manufacturers: 36 of 40 results within 5% of a Secondary Reference Laboratory; certificate valid 1 year, specific to reagents and instruments). ngsp.org/critsumm.asp
  6. NGSP. List of NGSP certified methods (updated monthly; certificates expire after 1 year). ngsp.org/certified.asp
  7. NGSP. CMS rule: from January 1, 2025, US HbA1c proficiency testing uses acceptable limits of plus or minus 8% (CAP GH-5: plus or minus 6%). ngsp.org/news.asp
  8. National Institute of Diabetes and Digestive and Kidney Diseases (NIH). The A1C Test & Diabetes. www.niddk.nih.gov/health-information/diagnostic-tests/a1c-test

Written and maintained by the controla1clevels.com editorial team. We do not have a medical reviewer; every claim links to its primary source above. Last reviewed: 2026-10-26. Spotted an error? See About.